Attenuation correction in quantitative SPECT of cerebral blood flow: a Monte Carlo study

Författare

  • A Arlig
  • A Gustafsson
  • L Jacobsson
  • Michael Ljungberg
  • C Wikkelso

Summary, in English

Monte Carlo simulation has been used to produce projections from a voxel-based brain phantom, simulating a 99mTc-HMPAO single photon emission computed tomography (SPECT) brain investigation. For comparison, projections free from the effects of attenuation and scattering were also simulated, giving ideal transaxial images after reconstruction. Three methods of attenuation correction were studied: (a) a pre-processing method, (b) a post-processing uniform method and (c) a post-processing non-uniform method using a density map. The accuracy of these methods was estimated by comparison of the reconstructed images with the ideal images using the normalized mean square error, NMSE, and quantitative values of the regional cerebral blood flow, rCBF. A minimum NMSE was achieved for the effective linear attenuation coefficient mu(eff) = 0.07 (0.09) cm(-1) for the uniform(pre) method, the effective mass attenuation coefficient mu(eff)/rho = 0.08 (0.10) cm2 g(-1) for the uniform(post) method and mu(eff)/rho = 0.12 (0.13) cm2 g(-1) for the non-uniform(post) method. Values in parentheses represent the case of dual-window scatter correction. The non-uniform(post) method performed better, as measured by the NMSE, both with and without scatter correction. Furthermore, the non-uniform(post) method gave, on average, more accurate rCBF values. Although the difference in rCBF accuracy was small between the various methods, the same method should be used for patient studies as for the reference material.

Avdelning/ar

  • Medicinsk strålningsfysik, Lund

Publiceringsår

2000

Språk

Engelska

Sidor

3847-3859

Publikation/Tidskrift/Serie

Physics in Medicine and Biology

Volym

45

Issue

12

Dokumenttyp

Artikel i tidskrift

Förlag

IOP Publishing

Ämne

  • Radiology, Nuclear Medicine and Medical Imaging

Status

Published

ISBN/ISSN/Övrigt

  • ISSN: 1361-6560