Ronnie Wirestam
Professor
Magnetization transfer (MT) of human brain at 7T in the context of a 3D multi-parameter mapping protocol
Författare
Summary, in English
mapping (MPM) of human brain. The increased polarization at 7T compared to lower B0 allows for increased spatial resolution or reduced scan
times. However, SAR restrictions imposed on the MT pulse and B1 inhomogeneity pose challenges. In this work, we propose a protocol for MPM of
human brain at 7T with special attention paid to eliminating bias when mapping MT saturation while obtaining sub-millimeter isotropic spatial resolution in
under 12 minutes with acceptable SNR.
Avdelning/ar
- Medicinsk strålningsfysik, Lund
- MR Physics
Publiceringsår
2019-04-29
Språk
Engelska
Publikation/Tidskrift/Serie
Proc Intl Soc Magn Reson Med
Volym
27
Länkar
Dokumenttyp
Konferensbidrag
Förlag
International Society for Magnetic Resonance in Medicine
Ämne
- Medical and Health Sciences
- Other Physics Topics
Conference name
ISMRM 27th Annual Meeting and Exhibition
Conference date
2019-05-11 - 2019-05-16
Conference place
Montreal, Canada
Aktiv
Published
Projekt
- Multiparametric mapping of the brain at 7T using gradient echoes
Forskningsgrupp
- MR Physics
ISBN/ISSN/Övrigt
- ISSN: 1545-4428